Black Friday Blowout!
Specification
Overview
Processor
Display
RAM
Storage
Video Card
Connectivity
Features
Battery
General
Description
Anesthesia Workstation
Features
lntegrated structure & Attractive DesignAccurate Manual Ventilation
Gas Supply: O2,N20,4 tube flowmeters(Optional: 6 tube flowmeters)Pneumatic system: Hypoxic guard system , O2 flush
Integrated Absorber system ,Autoclavable,Easy assembling & mountingVaporizer mount: Selectatec bar support 2 vaporizers
Vaporizer types: Enflurane, lsoflurane, Sevoflurane,Halothane
optional: Yoke system(N20,O2),AGSS),Suction,Auxiliary oxygen flowmeter and ACGO
Anesthesia Ventilator (Optional)
Suitable for infant Pediatric and adults
Tidal Volume setting 50 ~1500mL
User friendly interface,5.7″TFT LCD screen
Ventilation modes: vCV,Manual, Standby (Optional: PCV)Built-in active expiratory PEEP valve, built-in batteryo2 monitoring (Optional)
Measurement
Pressure values: Ppeak,Pmean,Pmin,Pplat,PEEPVolume/Flow values: Vti,Vte,MV,MVspn
FiO2 , f, fspn,Rst, Cdyn
Waveforms: Paw-t ,Flow-t
Alarms
MV, Paw, Continuous Paw high, Vte, Rate, FiO2, Mains failure, Battery low,Batterydischarged,O2 supply failure,Apnea.
| Power Specifications | Respiratory System | ||
| Parameter | Specification | Absorber Canister Capacity | About 1500mL |
| External | Connection | Inhalation/Exhalation ports;
Standard OD 22mm, ID 15mm, tapered connector; Manual breathing bladder port: Standard OD 22mm |
|
| Input Voltage | 100~240V | System leakage | In any mode, the system leak is less than 100ml/min |
| Input Frequency | 50/60Hz | System compliance | The loss due to the system’s internal compliance gas volume (manual mode) as follows: Type<=4mL/cmH2O |
| Input Power | <900 VA (with three AC outlets | Respiratory system resistance | Inspiratory: less than 0.6kPa
Expiratory: less than 0.6kPa |
| Internal Battery | Real-time clock | ||
| Battery Type | NiMH Batteries | Range | 2000 (00:00) ~ 2100 (23:59) |
| Battery voltage | 12 VDC | Accuracy | +-1 minute |
| Battery capacity | 4200mAh | Display Resolution | 1 minute/month
(at 21+-3 degrees under the conditions) |
| Min, Supply time | 90 min | Mode | |
| Charging time | 4h | Modes | VCV,PCV(option), Standby Manual |
| System Specificatoins | Flowmeter | Four-tube flowmeter, oxygen, nitrous oxide
Adjustment range Oxygen: 0.1 – 10 L / min (20 degrees) nitrous oxide: 0.1 – 12 L / min(20 degrees) Proportion of oxygen and nitrous oxide adjusted to ensure that the oxygen concentration more than 21% |
|
| Parameter | Specification | Auxiliary oxygen flowmeter (optional) | 0.1 – 15 L/min |
| Machine | Safety Pressure | System pressure does not exceed 12.5 kPa | |
| Size | 1410 * 950 * 650 (H*W*D) | Parameters Setting | |
| Weight | Approximately 110kg | Parameter | Specification |
| Max, load on top cover | 30kg | Tidal volume | Range: 20/50 – 1500mL;
Increment: 20/50 – 100 mL; 5mL 100 – 1500 mL; 10mL |
| Display | Respiratory rate | Range: 1 – 100 bpm
Increment: 1 bpm |
|
| Type | Color TFT LCD (touch screen) | Inspiratory time | Range: 0.1 – 10.0 s; increments: 0.1 s |
| Size | 5.7 inches | Respiratory ratio | Range: 4:1 to 1:10; increments: 0.5s |
| Resolution | 640 * 480 pixels | Percentage of inspiratory pause | Range: 0 to 50%; Increment: 5% |
| Brightness | Not Adjustable | PEEP | Range: OFF, 3 – 30 cmH2)
Increment: 1 cmH2O |
| LED Indication | Pressure Support | Range: 0 – 70 cmH2O
Increment: 1 cmH2O |
|
| Alarm Indication | 8 (yellow, red, and when the senior and mid-level alarms occur simultaneously, only flashed red) | Pressure Control | Range: 5 – 70 cmH2O
Increment: 1 cmH2O |
| AC Power Indicator | 1 (green) | Flow trigger | Range: 0.5 – 20L/min;
Increments: 0.1 L/min |
| Battery Indicator | 2 (green and orange) | Pressure Trigger | Range: 0 20 cmH2O
Increments: 0.1 cmH2O |
| Audio Instruction | Monitoring | ||
| Speaker | Alarm sounds, tone: supports multi-level volume, alarm tone meet IEC60601_1_8 standards | Parameter | Description |
| Control | Inspiratory tidal volume | Range: 0 – 2500 mL | |
| Knobs | 1
Support Clockwise / counter-clockwise rotation and pressing operation |
Expiratory tidal volume | Range: 0 – 2500 mL |
| Button | Minute volume | Range: 0 – 60 L/min | |
| Interface | Spont minute volume | Range: 0 – 60 L/min | |
| Power supply | An AC power connector
Three auxiliary output power interface |
Respiratory rate | Range: 0 – 100 L/min |
| Equipotential | An equipotential ground | Spontaneous breathing frequency | Range: 0 – 100 L/min |
| USB | 1 standard USB interface | Respiratory ratio | Range: 30:1 to 1:150 |
| Drawer | Drawer: 200*392*398 (H*W*D)
Drawer two 200*392*398 (H*W*D) |
Peak airway pressure | Range: 0 – 100 L/min |
| Respiratory System | Mean airway pressure | Range: 0 – 100 L/min | |
| Absorber Canister Capacity | About 1500mL | PEEP | Range: 0 – 100 L/min |
| Inspiratory plateau pressure | Range: 0 – 100 L/min | ||
| FiO2(optional) | Range: 15 to 100% | ||
| Compliance | Range: 0 – 300 mL/cmH2O | ||
| Airway resistance | Range: 0 – 600 cmH2O / (L/S) | ||


Reviews